The .gov means it’s official. Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site. The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely. As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsement of, or agreement with, the contents by NLM or the National Institutes of Health. Learn more about our disclaimer.
Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2018 Aug; 35(4): 539–549.
PMCID: PMC9935132

Language: Chinese | English

基于递归最小二乘法的回声状态网络算法用于心电信号降噪

An echo state network algorithm based on recursive least square for electrocardiogram denoising

杰烁 张

河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China

Find articles by 杰烁 张

明 刘

河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China

Find articles by 明 刘

鑫 李

河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China

Find articles by 鑫 李

鹏 熊

河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China

Find articles by 鹏 熊

秀玲 刘

河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China 河北大学 电子信息工程学院 河北省数字医疗工程重点实验室(河北保定 071002), Key Laboratory of Digital Medical Engineering of Hebei Province, College of Electronic and Information Engineering, Hebei University, Baoding, Hebei 071002, P.R.China

刘明,Email: moc.621@reelg

目前,用于 ECG 信号降噪的方法主要有:人工神经网络(artificial neural networks,ANN)、集合经验模式分解(ensemble empirical mode decomposition,EEMD)、主成分分析(principal component analysis,PCA)、独立成分分析(independent component analysis,ICA)、傅里叶变换、S 变换和小波变换等 [ 3 - 11 ] 。Jebaraj 等 [ 4 ] 利用 EEMD 将含噪 ECG 信号分解为若干个不同时间尺度的固有模态函数(intrinsic mode function,IMF),然后将这些 IMF 中含有噪声的舍弃,最终能够滤除含噪信号中 60 Hz 的电力线干扰。但是,在舍弃的 IMF 中也会含有与低频噪声的频率相似的 ECG 的 P 波和 T 波,这就导致重构的 ECG 信号不完整。而 PCA 和 ICA 都是用于从非高斯噪声信号中寻找独立的源信号的盲源分离技术,能够滤除与 ECG 信号统计不相关的噪声 [ 5 - 6 ] 。但是,该类方法的鲁棒性低,对于含噪 ECG 信号的细微变化十分敏感,并不适用于远程医疗背景下 ECG 信号中含有复杂噪声的情况。Alyasseri 等 [ 11 ] 首先利用 β 爬山算法选择的最优的小波基将含噪 ECG 信号由时域变换到频域的不同尺度中,再利用阈值保留主要由有效信号控制的小波系数,最终该方法在去除高频大噪声方面取得了不错的结果。但对于 ECG 信号和噪声的频带相互重叠的情况,主要保留的小波系数仍然含有部分噪声信息,这也就必然导致重构的 ECG 信号的低频成分存在形态失真。基于以上分析,现有研究成果在降噪精度上均得到了较满意的效果,但并没有过多地关注于 ECG 信号的低频成分保持问题。

因此,为了进一步提高在远程医疗背景下被复杂噪声污染的 ECG 信号的降噪效果,尤其是对 ECG 信号低频成分的保持,本文提出了基于递归最小二乘(recursive least square,RLS)的回声状态网络(echo state network,ESN)算法——RLS-ESN 算法。该方法利用 RLS 来训练 ESN,从而使网络自动学习采集到的含噪 ECG 数据中非线性的且具有区分度的深层次特征,最后网络利用这些能够区分 ECG 信号和噪声的特征自动重构得到干净的 ECG 信号。

1. 方法

图 1 所示,基于 RLS 在线训练算法的 ESN 降噪框图主要由 4 部分组成:预处理信号、初始化 ESN、构建输入输出函数和基于 RLS 的在线训练网络。首先,由于过大的采样频率会造成 ECG 信号中冗余信息增多降低网络运行效率,而过小的采样频率又会丢失 ECG 信号中有效信息,因此本文算法对 ECG 信号以合适的 250 Hz 的采样频率进行重采样;其次,依据网络参数的范围设置合适的 ESN 网络参数;然后,为了更符合现实需求和 ECG 信号非平稳信号的特点,采用了逐个处理信号的 RLS 在线训练算法对 ESN 进行训练,利用储备池的稀疏性能得到含噪 ECG 数据中非线性的且具有区分度的深层次特征,并且通过训练得到最优的输出权值矩阵,保存训练好的 ESN 网络;最后,将测试集中含噪 ECG 信号输入训练好的 ESN 网络,利用储备池得到能够区分 ECG 信号和噪声的特征后,训练好的 ESN 网络再利用输出权值矩阵重构得到干净的 ECG 信号。

ECG signal filter block diagram of ESN with RLS online training algorithm

基于 RLS 在线训练算法的 ESN 降噪框图

1.1. 数据来源

本文实验中使用的 ECG 信号来自美国麻省理工学院与贝斯以色列医院(Massachusetts Institute of Technology and Beth Israel Hospital,MIT-BIH)联合构建的数据库中的心律失常数据库(Arrhythmia Database)(网址为: https://www.physionet.org/physiobank/database/mitdb/ ),该数据库中包括 48 组采样频率为 360 Hz 并且数据长度为 30 min 的动态 ECG 信号记录,每组记录包括来自医用的 24 h 动态 ECG 信号记录仪的 12 导联数据中的两个导联信号:第一个导联信号为修正的肢体Ⅱ导联,第二个导联信号为胸部 V2 导联或胸部 V5 导联 [ 12 ] 。为了与已有文献中的采用基于子带自适应阈值的小波变换(wavelet transform with subband dependent threshold,WT-Subband)降噪算法和 S 变换降噪算法进行比较,本文对应展示了 9 个 ECG 数据的实验结果,分别为:103 号、105 号、111 号、116 号、122 号、205 号、213 号、219 号、223 号。

实验中采用的噪声信号来自于 MIT-BIH 数据库中的噪声压力数据库(Noise Stress Test Database)(网址为: https://www.physionet.org/physiobank/database/nstdb/ ),其中包括 3 h 的动态 ECG 信号记录噪声数据,分别为基线漂移(baseline wander,BW)、肌电干扰(muscle artifacts,MA)和电极干扰(electrode motion,EM)共 3 种噪声数据。

1.2. 基于 RLS-ESN 的 ECG 信号降噪算法

ESN 作为深度学习的主流模型之一,不但具有基于医疗大数据背景下数据增多效果更优的特点,而且还具有求解过程简单且训练速度快的特点。它的独特之处在于其是采用由随机稀疏连接的神经元组成的储备池来代替传统神经网络中的隐藏层,并且唯一需要训练的参数只有输出权值矩阵,因此大大简化了网络权值的求解过程和训练算法 [ 13 - 14 ] 。基于 ESN 的降噪算法是通过训练网络参数自动学习得到含噪 ECG 数据中非线性的且具有区分度的深层次特征,并利用这些特征以实现将 ECG 信号与噪声分离的目的。

本文提出了基于 RLS 的在线训练算法完成 ESN 降噪模型的权重更新。相比于离线训练算法一次性的处理所有的输入信号,以序列化的方式逐个处理输入信号的基于 RLS 的在线算法更适合于目前信息处理的需求 [ 15 ] 。并且 RLS 的基本思想是在每个时刻对所有已输入信号而言重估的平方误差的加权和最小,这使得其对 ECG 信号这种非平稳信号的适应性非常好 [ 16 ] 。基于 RLS 在线训练算法的 ESN 降噪模型的结构图如 图 2 所示。

The structure of ESN based on RLS online training algorithm

基于 RLS 在线训练算法的 ESN 结构图

1.2.1. 信号预处理

作为一种准周期性信号,ECG 信号有效信息主要分布在 P-QRS-T 波群中,冗余信息主要体现在相邻样本点之间的相关性和相邻心拍之间的相似性 [ 17 ] 。因此,在处理同等时间长度的 ECG 信号时,过大的采样频率会导致数据冗余增多,并且冗余信息的增多不仅没有为 ESN 降噪模型增添新的特征,反而会增大储备池维度,降低 ESN 的计算速率增大计算误差导致网络运行性能降低;而过小的采样频率则直接会导致 ECG 信号中相关性等特征信息的丢失。

基于此,Petrėnas 等 [ 18 ] 根据人的心理电活动周期并且经过多次试验得出,当 ECG 信号的频率为 250 Hz 时,可以在包含 ECG 信号主要特征的前提下使用较低纬度的储备池,以保证 ESN 的高效运行。因此本文预处理时,首先将 MIT-BIH 库中的 48 组 ECG 数据以及 3 种噪声数据重采样为 250 Hz。然后,将 3 种噪声数据按照一定的信噪比(signal-to-noise ratio,SNR)(以符号 SNR 表示)叠加到心律失常数据库中的 48 组 ECG 信号上,即可得到含噪 ECG 信号。最后,利用每个原始 ECG 数据的前 3 min 信号和相应的含噪 ECG 数据的前 3 min 信号组成训练集,剩余的含噪 ECG 信号组成测试集。

1.2.2. ESN 降噪模型的初始化

为消除 ECG 信号中的噪声,需构建基于 RLS 在线训练算法的 ESN 降噪模型。首先,进行参数的初始化。ESN 的主要参数包括:储备池规模、储备池稀疏程度、储备池谱半径和储备池输入单元尺度 [ 19 ] 。储备池规模(符号记为: N )定义为储备池中的神经元的个数,在储备池内部通过这 N 个神经元之间稀疏连接形成 N × N 的储备池内部神经元连接权值矩阵 W 。储备池稀疏程度(符号记为: SD )定义为储备池中相互连接的神经元个数与总神经元个数之比,表达式为 SD = m / N ,通常 SD 的取值为 1%~5%。储备池谱半径(符号记为: SR )定义为 W 的最大特征值的绝对值,表达式为 SR = max{abs( W 的特征值)},且当 SR < 1 时,ESN 才具有回声状态性质。储备池输入单元尺度(符号记为: IS )是一个尺度因子,在输入信号连接至储备池之前与输入信号相乘,将输入信号变换至神经元激活函数相应的范围内。

其次,进行权值矩阵的初始化。输入单元与储备池内部连接的的权值矩阵 W in 和储备池内部神经元连接权值矩阵 W 的初始化都是随机且服从均匀分布的,且初始给定后固定不变。而储备池与输出单元连接权值矩阵 W out 和输出单元与储备池的连接权值矩阵 W back 初始化为零矩阵,其中 W out 通过训练得到其最终值。

1.2.3. 输入输出信号的构建

ESN 的输入信号 x r n )的表达式如式(1)所示。其中, x n )为当前时刻的含噪 ECG 信号; x n -1)为上一时刻的含噪 ECG 信号; x ′( n )和 x ″( n )分别为 x n )的一阶导和二阶导,表征了 x n )的细节特征。 ,其中 r n )是储备池的状态矩阵。

误差定义为当前时刻的干净 ECG 信号 d n )与 y n )的差值,表达式为 e n )= d n )– y n ),其中 d n )为当前时刻的干净 ECG 信号。因此,在训练中需要通过更新 W out r n )使滤波后的 ECG 信号 y n )逼近干净 ECG 信号 d n ),使误差 e n )能够近似于 0。

1.2.4. 基于 RLS 的在线训练

在本文中,输出权重 W out 的更新选用基于 RLS 的在线训练算法 [ 20 ] 。具体过程如下:

(1) u n )的表达式为 ,其中 ψ n )为 v n )的相关系数矩阵的逆矩阵,并且 ψ n )的更新如式(3)、式(4)所示:

2.2. 实验结果与讨论

本文经过多次实验后,网络参数初始化为 SR = 0.4, SD = 2%, λ = 0.999 9, δ = 0.001 72, α = 0.8, IS = [0.01;0.01;0.9;0.8],输入神经元个数为 4,储备池规模 N 为 1 000,输出神经元个数为 1。

2.2.1. 实验结果

滤除 1.25 dB 基线漂移的实验结果如 图 3 所示。其中,含噪 ECG 信号和原始 ECG 信号(以 MIT-BIH 数据库中 103 号为例)相比较可以看出,基线漂移的存在已经影响了整体波形的走势;并且由含噪信号的局部细节图可以看出,基线漂移使得正常的 ST 段抬高成为凸面向上型。但经过本文提出的基于 RLS 在线训练算法的 ESN 降噪算法降噪后的 ECG 信号和原始 ECG 信号相比较可以看出,降噪后的信号回到了基线,且形态和原始信号基本保持一致,尤其从降噪后的 ECG 信号的局部细节图可以发现,ECG 信号的 ST 段已恢复为正常的波形。

An external file that holds a picture, illustration, etc. Object name is swyxgcxzz-35-4-539-3.jpg

Experimental results for baseline wander noise case

滤除基线漂移结果图

滤除 1.25 dB 肌电干扰的实验结果如 图 4 所示。其中,含噪 ECG 信号和原始 ECG 信号(以 103 号为例)相比较可以看出,肌电干扰已严重破坏了 ECG 信号中特征波的形态;并且由含噪 ECG 信号的局部细节图可以发现,Q 波和 S 波、P 波、T 波等低频成分已经被肌电干扰完全淹没,仅剩下了幅值较大的 R 波,造成了 ECG 信号的低频特征丢失。但经过本文提出的基于 RLS 在线训练算法的 ESN 降噪算法降噪后的 ECG 信号和原始 ECG 信号相比较可以看出,本文的降噪算法在滤除肌电干扰时,完整地保留了原有波形的形态特征,尤其从降噪后的 ECG 信号的局部细节图可以发现,ECG 信号的低频波形特征也得到了很好的保留。

An external file that holds a picture, illustration, etc. Object name is swyxgcxzz-35-4-539-4.jpg

Experimental results for muscle artifacts noise case

滤除肌电干扰结果图

滤除 1.25 dB 电极干扰的实验结果如 图 5 所示。其中,含噪 ECG 信号和原始 ECG 信号(以 103 号为例)相比较可以看出,电极干扰已经改变了原始 ECG 信号中特征波的形态,部分噪声甚至在特征波未出现的地方模仿出了特征波的形态;并且由含噪 ECG 信号的局部细节图可以发现,电极干扰不但更改了 P 波和 T 波的形态,更是模拟出了原本没有的 U 波,使得正常的 T 波变成了 U 波增大型的 T 波。但经过本文提出的基于 RLS 在线训练算法的 ESN 降噪算法降噪后的 ECG 信号和原始 ECG 信号相比较可以看出,本文的降噪算法很好地滤除了 ECG 信号中电极干扰,恢复了信号原有的形态特征,尤其从降噪后的 ECG 信号的局部细节图可以发现,该降噪算法滤除了噪声模拟出的伪 U 波,并对于 ECG 信号的 P 波和 T 波等微弱波形特征也有较高的复原度。

An external file that holds a picture, illustration, etc. Object name is swyxgcxzz-35-4-539-5.jpg

Experimental results for electrode motion noise case

滤除电极干扰结果图

图 3 图 5 所示,3 种不同的噪声(基线漂移、肌电干扰、电极干扰)都对 ECG 的形态产生了破坏,出现一些伪病态波形,比如基线改变、P 波消失或者伪 U 波出现等。而从每幅图的原始信号和处理后的信号对比可以看出,本文提出的降噪算法很好地滤除了噪声,并且由每幅图降噪后的 ECG 信号局部细节图可以发现,降噪后的 ECG 信号很好地恢复了原始信号的形态特征。对于一些微弱低频成分,例如:P 波和 T 波,由于其幅值小频率低,一方面导致其极易被噪声淹没或改变形态,另一方面也使得其在降噪过程中容易被当成噪声滤除,但是在本文方法的降噪结果中,这些幅值很小的低频细节特征也得到了很好的恢复和保持。

2.2.2. 与其他降噪方法的比较

表 1 所示,分别为 WT-Subband 法、S 变换法和本文所提出的基于 RLS 在线训练算法的 ESN 降噪算法来滤除 0 dB、1.25 dB 及 5 dB 的基线漂移的实验结果,选取 MIT-BIH 数据库中共 9 个 ECG 数据的实验结果,分别为:103 号、105 号、111 号、116 号、122 号、205 号、213 号、219 号、223 号。

表 1

Experimental results for baseline wander noise

滤除基线漂移的实验结果

数据 WT-Subband 法 S 变换法 本文方法
0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB
SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB
103 0.16 0.981 1.41 0.849 5.16 0.551 11.40 0.269 12.06 0.249 13.54 0.210 17.72 0.041 18.23 0.039 19.82 0.032
105 0.20 0.977 1.45 0.846 5.20 0.549 11.56 0.264 12.23 0.245 13.77 0.205 15.08 0.053 15.48 0.051 16.49 0.045
111 0.21 0.974 1.47 0.842 5.21 0.546 9.22 0.346 9.61 0.331 10.41 0.302 12.90 0.041 13.24 0.039 14.12 0.035
116 0.33 0.962 1.49 0.842 5.22 0.548 9.01 0.354 9.35 0.341 10.04 0.315 14.87 0.108 15.99 0.095 16.21 0.092
122 0.33 0.962 1.60 0.830 5.42 0.534 10.58 0.296 11.17 0.276 12.38 0.240 16.80 0.050 17.13 0.048 19.20 0.038
205 0.51 0.942 1.82 0.810 5.73 0.515 9.31 0.342 9.70 0.327 10.46 0.300 17.97 0.021 18.77 0.019 20.16 0.017
213 0.22 0.976 1.46 0.844 5.22 0.546 11.57 0.264 12.22 0.245 13.77 0.205 14.54 0.104 15.29 0.096 15.78 0.090
219 0.21 0.976 1.48 0.843 5.29 0.544 11.55 0.265 12.22 0.245 13.65 0.208 10.84 0.133 12.00 0.116 12.67 0.107
223 0.27 0.969 1.53 0.838 5.32 0.542 12.14 0.247 12.95 0.225 14.81 0.182 14.68 0.073 14.59 0.074 15.10 0.069

图 6 所示,以柱状图的形式展示了三种方法滤除 1.25 dB 噪声后的结果。相比于使用前两种方法降噪,使用本文提出的方法降噪后 SNR 和 RMSE 与前两种方法相比有了明显的改善,表明了本文所提出的算法可以有效地滤除基线漂移噪声。例如,122 号数据叠加 1.25 dB 的基线漂移噪声后,其经过 WT-Subband 法和 S 变换法降噪后的 SNR 值分别为 1.60 dB 和 11.17 dB,RMSE 值分别为 0.830 和 0.276;经本文提出的算法降噪后的 SNR 值达到了 17.13 dB,RMSE 值降为了 0.048。

De-noised results for 1.25 dB baseline wander noise case

滤除 1.25 dB 基线漂移的结果对比图

其中 219 号数据的实验结果中,本文所提出的方法与 S 变换法处理后的 SNR 值相差不多,RMSE 值则有相当大的提高,这是由于 219 号数据患者本身患有阵发性房颤,这导致其 ECG 信号中阵发性出现特异性心拍,且 RR 间期不正常。但其训练集中出现特异性心拍的个数较少,测试集中较多。本文所提出的降噪算法是依据训练集训练后所保存的 ESN 网络来进行降噪处理,当输入与训练样本相差较大的数据时其实验结果也会较差,同时如果其训练样本增多,其实验结果也将更优。

表 2 所示,分别为采用 WT-Subband 法、S 变换法和本文所提出的基于 RLS 在线训练算法的 ESN 降噪算法对含有肌电干扰的 ECG 信号进行滤波后的结果,并以柱状图的形式展示了这三种方法分别滤除 1.25 dB 噪声后的 SNR 值和 RMSE 值,如 图 7 所示。相比于使用前两种方法降噪,使用本文提出的方法降噪后 SNR 值和 RMSE 值与前两种方法相比有了一定的改善,表明了本文提出的降噪算法可以有效地滤除肌电干扰。例如,122 号数据叠加 1.25 dB 的肌电干扰后,其经过 WT-Subband 法和 S 变换法降噪后的 SNR 值分别为 2.86 dB 和 9.67 dB,RMSE 值分别为 0.729 和 0.328;经本文提出的算法降噪后的 SNR 值达到了 14.27 dB,RMSE 值降为了 0.069。

表 2

Experimental results for muscle artifacts noise

滤除肌电干扰的实验结果

数据 WT-Subband 法 S 变换法 本文方法
0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB
SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB
103 1.62 0.838 2.82 0.730 6.40 0.482 10.41 0.302 10.89 0.286 12.63 0.234 10.51 0.099 13.11 0.070 14.15 0.065
105 1.67 0.836 2.90 0.726 6.49 0.479 10.02 0.316 10.42 0.301 12.76 0.230 10.40 0.085 11.47 0.075 13.79 0.057
111 1.49 0.850 2.70 0.739 6.17 0.494 8.21 0.389 8.66 0.369 9.94 0.318 10.16 0.064 11.40 0.056 13.02 0.046
116 1.55 0.847 2.76 0.736 6.29 0.489 8.19 0.389 8.51 0.375 9.75 0.325 9.37 0.203 10.53 0.178 13.45 0.127
122 1.64 0.839 2.86 0.729 6.49 0.478 9.20 0.347 9.67 0.328 11.69 0.260 12.56 0.084 14.27 0.069 15.01 0.063
205 1.72 0.828 2.95 0.718 6.54 0.473 8.32 0.384 8.61 0.371 9.91 0.320 11.24 0.049 12.24 0.043 13.97 0.036
213 1.42 0.856 2.58 0.748 6.02 0.501 8.79 0.363 9.67 0.329 12.53 0.236 9.29 0.196 10.46 0.171 12.34 0.138
219 1.64 0.838 2.85 0.728 6.44 0.481 10.05 0.314 10.61 0.295 12.89 0.227 9.47 0.140 9.85 0.149 12.62 0.097
223 1.68 0.836 2.92 0.726 6.56 0.477 9.95 0.318 10.56 0.297 13.44 0.213 11.69 0.099 13.05 0.084 14.82 0.069

De-noised results for 1.25 dB muscle artifacts noise case

滤除 1.25 dB 肌电干扰的结果对比图

表 3 所示,分别为采用 WT-Subband 法、S 变换法和本文所提出的基于 RLS 在线训练算法的 ESN 降噪模型算法对含有电极干扰的 ECG 信号进行滤波后的实验结果,如 图 8 所示,为三种方法分别滤除 1.25 dB 电极干扰后的 SNR 值和 RMSE 值。相比于使用前两种方法降噪,使用本文提出的方法降噪后 SNR 值和 RMSE 值与前两种方法相比有了很大的改善。例如,122 号数据叠加 1.25 dB 的电极干扰后,其经过 WT-Subband 法和 S 变换法降噪后的 SNR 值分别为 1.75 dB 和 6.96 dB,RMSE 值分别为 0.816 和 0.449;经本文提出的算法降噪后的 SNR 值达到了 12.48 dB,RMSE 值降为了 0.083。

表 3

Experimental results for electrode motion noise

滤除电极干扰的实验结果

数据 WT-Subband 法 S 变换法 本文方法
0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB 0 dB 1.25 dB 5 dB
SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB SNR /dB
103 0.26 0.941 1.21 0.837 5.29 0.543 6.41 0.478 7.47 0.423 10.32 0.305 13.19 0.069 13.83 0.064 15.47 0.053
105 0.27 0.934 1.58 0.833 5.30 0.542 6.13 0.493 7.35 0.429 10.40 0.302 10.31 0.092 11.13 0.084 12.88 0.069
111 0.30 0.935 1.56 0.834 5.30 0.540 5.45 0.534 6.40 0.478 8.54 0.374 9.03 0.064 9.50 0.060 11.01 0.051
116 0.41 0.944 1.66 0.825 5.42 0.535 5.47 0.533 6.32 0.483 8.32 0.384 12.08 0.149 12.88 0.135 14.21 0.116
122 0.01 0.906 1.75 0.816 5.53 0.528 5.87 0.509 6.96 0.449 9.60 0.331 11.76 0.090 12.48 0.083 14.25 0.067
205 0.57 0.935 1.87 0.805 5.76 0.513 5.59 0.525 6.47 0.475 8.55 0.374 14.32 0.033 14.74 0.031 15.94 0.027
213 0.37 0.947 1.60 0.830 5.26 0.543 5.85 0.510 7.06 0.444 10.12 0.312 10.82 0.160 11.16 0.154 12.78 0.128
219 0.13 0.903 1.61 0.831 5.39 0.537 6.05 0.498 7.17 0.438 10.04 0.315 8.01 0.184 9.10 0.162 10.34 0.141
223 0.13 0.862 1.62 0.830 5.42 0.535 6.21 0.489 7.45 0.424 10.74 0.290 10.11 0.123 10.84 0.113 12.28 0.096

De-noised results for 1.25 dB electrode motion noise case

滤除 1.25 dB 电极干扰的结果对比图

表 1 表 3 的这 3 个表中以及 图 6 图 8 的三个柱状图中可以看出,与 WT-Subband 法和 S 变换法相比,本文基于 RLS 在线训练算法的 ESN 降噪模型在去除 ECG 信号中的大噪声时,无论是反映降噪效果总体误差的 RMSE 值的降低,还是反映降噪后信号和噪声的比值关系的 SNR 值的较大提高,都验证了本文提出的基于 RLS 在线训练算法的 ESN 降噪模型对 ECG 信号中的三种复杂噪声有着较好的降噪效果。

3. 结论

针对消除 ECG 信号包含的复杂噪声并完整保留 ECG 信号的形态这一难题,本文提出了基于 RLS 在线训练算法的 ESN 降噪模型。与 WT-Subband 法和 S 变换法的实验结果对比表明,基于 RLS 在线训练算法的 ESN 降噪模型的实验结果在 SNR 值和 RMSE 值上都具有明显优势,能够达到较好的降噪效果。并且,采用 MIT-BIH 数据库中的数据进行对比实验结果表明,该降噪模型在处理被噪声污染较严重的 ECG 信号时,不仅有较好的降噪效果,而且 ECG 信号的低频特征波也得到了很好的恢复和保持。综上所述,本文实验结果表明,本文所提出的降噪模型在心血管疾病的智能诊断的预处理阶段具有很好的应用价值和实际意义。

Funding Statement

国家自然科学基金资助项目(61673158);国家自然科学基金青年基金资助项目(61703133);河北省杰出青年科学基金资助项目(F2016201186);河北省研究生创新资助项目(CXZZSS2017010);河北省高等学校技术研究重点项目(ZD2015067);保定市科技计划项目(17H01)

References

1. 陈伟伟, 高润霖, 刘力生, 等 《中国心血管病报告 2017》概要 中国循环杂志 2018; 33 (1):1–8. [ Google Scholar ]
2. 刘胜洋, 张根选, 曹阳 心电信号中的工频干扰滤除方法 生物医学工程学杂志 2014;(3):577–582. [ Google Scholar ]
3. Kaergaard K, Jensen S H, Puthusserypady S A comprehensive performance analysis of EEMD-BLMS and DWT-NN hybrid algorithms for ECG denoising. Biomed Signal Process Control. 2016; 25 :178–187. [ Google Scholar ]
4. Jebaraj J, Arumugam R Ensemble empirical mode decomposition-based optimised power line interference removal algorithm for electrocardiogram signal. IET Signal Processing. 2016; 10 (6):583–591. [ Google Scholar ]
5. Alickovic E, Subasi A Effect of multiscale PCA de-noising in ECG beat classification for diagnosis of cardiovascular diseases. Circuits Systems and Signal Processing. 2015; 34 (2):513–533. [ Google Scholar ]
6. Barhatte A S, Ghongade R, Tekale S V. Noise analysis of ECG signal using fast ICA// Advances in Signal Processing. Pune, 2016: 118-122
7. Sadhukhan D, Mitra M. ECG noise reduction using Fourier coefficient suppression// International Conference on Control, Instrumentation, Energy and Communication. Calcutta, 2014: 142-146
8. Ari S, Das M K, Chacko A ECG signal enhancement using S-Transform. Comput Biol Med. 2013; 43 (6):649–660. [ PubMed ] [ Google Scholar ]
9. 刘秀玲, 乔磊, 杨建利, 等 改进小波阈值法用于心电信号去噪 生物医学工程学杂志 2014;(3):511–515. [ Google Scholar ]
10. Singh O, Sunkaria R K ECG signal denoising via empirical wavelet transform. Australasian Physical & Engineering Sciences in Medicine. 2017; 40 (1):219–229. [ PubMed ] [ Google Scholar ]
11. Alyasseri Z A, Khader A T, Al-Betar M A, et al Hybridizing beta-hill climbing with wavelet transform for denoising ECG signals. Inf Sci (Ny) 2018; 429 :229–246. [ Google Scholar ]
12. Goldberger A L, Amaral L A, Glass L, et al PhysioBank, PhysioToolkit, and PhysioNet: components of a new research resource for complex physiologic signals. Circulation. 2000; 101 (23):E215–E220. [ PubMed ] [ Google Scholar ]
13. Jaeger H. The “echo state” approach to analysing and training recurrent neural networks, überwachtes lernen, 2001:1-47
14. 韩敏, 孙磊磊, 洪晓军 基于回声状态网络的脑电信号特征提取 生物医学工程学杂志 2012; 29 (2):206–211. [ Google Scholar ]
15. Jaeger H, Haas H Harnessing nonlinearity:predicting chaotic systems and saving energy in wireless telecommunication. Science. 2004; 304 (5667):78–80. [ PubMed ] [ Google Scholar ]
16. 敖露 LMS和RLS算法在自适应逆控制中的应用 工业控制计算机 2012; 25 (5):75–77. [ Google Scholar ]
17. Zhang Bo, Zhao Jiasheng, Chen Xiao, et al ECG data compression using a neural network model based on multi-objective optimization. PLoS One. 2017; 12 (10):e0182500. [ PMC free article ] [ PubMed ] [ Google Scholar ]
18. Petrėnas A, Sörnmo L, Lukoševičius A, et al Detection of occult paroxysmal atrial fibrillation. Med Biol Eng Comput. 2015; 53 (4):287–297. [ PubMed ] [ Google Scholar ]
19. 彭宇, 王建民, 彭喜元 基于回声状态网络的时间序列预测方法研究 电子学报 2010; 38 (S1):148–154. [ Google Scholar ]
20. Park B G, Kim R Y, Hyun D S. Fault diagnosis using recursive least square algorithm for permanent magnet synchronous motor drives// IEEE International Conference on Power Electronics and Ecce Asia. Jeju, South Korea: 2011: 2506-2510

Articles from Sheng Wu Yi Xue Gong Cheng Xue Za Zhi = Journal of Biomedical Engineering are provided here courtesy of West China Hospital of Sichuan University